ADFC8MD Off-line Dissolution system is consists of RC8MD Dissiolution Tester,RZQ-8D Sampling and Collection System.Automatically complete dissolution and sampling process with rapid  on high precision.

 

Features

RC8MD  Dissolution Tester 

● Paddle wobble: 0.5 mm

● Basket wobble: 1.0 mm

● Axial deviation between rotating rod and dissolution vessel: 2.0 mm

● Position deviation of mensuration: ±2.0 mm

● Range of speed regulation:25250rpm

● Speed resolution ratio: 1 rpm 

●Stabilizing speed error:  ±1 rpm

● Range of temperature regulation:5.045.0 (room temperature)

● Temperature resolution: 0.1 

● Temperature control error:  ±0.3 

● Program-control test procedure: ≤10 groups

● Program-control test procedure: 10 groups

● Program-control sampling times: ≤12 times

● Program-control sampling period: ≤ 99 hrs 59 min

● Print cycle: ≤60 min

● Operation noise: 55dB

● Ambient conditions: Ambient temperature (5~35), RH≤80%

● Safety performance: in accordance with national industrial standards.

● Requirement for Power supply: AC 220V 50Hz 4A

● Dimension:650×570×360 mm

● Weight: approximately 50 kg

 

RZQ-8D  Sampling and collection System

 

1. Number of sampling access: 8

2. Maximum number of sampling: 15

3. Maximum interval: 99 hours and 59minutes

4. Minimum interval: 1 minute for the first interval, and 4~6 minutes for other intervals (depending on sampling dose and necessity for fluid replacement)

5. Continuous operating time: 99 hours 59 minutes×15

6. Sample filtering cycle: 30

7. Sampling doseor amount of fluid replacement):1ml10ml

8. Sampling accuracy: ±10%0.1ml

9. Bore diameter of filtration: columnar Teflon filter with a bore diameter of 10μm for primary filtration; injection filter with a maximum bore diameter of 0.22μm for secondary filtration.

10. Number of sampling access: 8

11. Maximum number of sampling: 15

12. Maximum interval: 99 hours and 59minutes

13. Minimum interval: 1 minute for the first interval, and 4~6 minutes for other intervals (depending on sampling dose and necessity for fluid replacement)

14. Continuous operating time: 99 hours 59 minutes×15

15. Sample filtering cycle: 30

16. Sampling doseor amount of fluid replacement):1ml10ml

17. Sampling accuracy: ±10%0.1ml

18. Bore diameter of filtration: columnar Teflon filter with a bore diameter of 10μm for primary filtration; injection filter with a maximum bore diameter of 0.22μm for secondary filtration.

 

 